Affiliation:
1. Upper Limb Unit, Department of Orthopaedics, Wrightington Hospital, Lancashire, UK
Abstract
Background Restoring the radial head as a stabiliser of the elbow is considered important in cases where there is an associated bony or ligamentous injury. A variety of radial head prostheses are available. There are no reports of the survivorship of a short-stem press-fit radial head prostheses with pyrocarbon bearing. Patients and methods With institutional review board approval, a retrospective case-note analysis was performed of a consecutive single-surgeon case series of the Integra LifeScience pyrocarbon radial head prostheses (Carbon Modular Radial Head) from October 2010 to October 2019 in a tertiary referral centre. The series was divided into acute trauma and salvage cohorts. Kaplan–Meier survivorship analysis was conducted. Results 36 patients were included, 23 (64%) for acute injuries and 13 (36%) for failed initial treatment. Of the acute injuries, 20 (87%) had a Wrightington type-C elbow fracture-dislocation. Reoperation was performed in 4 (11%) patients; 10 (28%) had possible loosening on radiographs. The implant survival rate was 94% at shortest follow-up of 17 months (mean 70 months). Discussion Treatment of complex radial head fractures using a pyrocarbon-bearing, anatomic press-fit design provides satisfactory short-term survivorship in this case series. The implant should be used with caution in salvage cases, due to higher rates of loosening seen in this cohort.
Subject
Rehabilitation,Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine,Surgery
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献