Author:
Flood-Joy Maureen,Mathiowetz Virgil
Abstract
These two studies evaluated whether three types of Jamar dynamometers measured equivalently. In study A, 26 subjects were tested on 2 types; in study B, 30 subjects were tested on 3 types. Both studies found significant differences between and among the dynamometers. Therapists can no longer assume that all versions of the Jamar dynamometer measure equivalently, and the differences should be taken into account when using normative data and when assessing treatment effectiveness. In the latter case, the same dynamometer must be used for pre- and posttesting.
Cited by
35 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献