Optical Coherence Tomography of the Tympanic Membrane and Middle Ear: A Review

Author:

Tan Hsern Ern Ivan123,Santa Maria Peter Luke124,Wijesinghe Philip56,Francis Kennedy Brendan56,Allardyce Benjamin James7,Eikelboom Robert Henry128,Atlas Marcus David12,Dilley Rodney James12

Affiliation:

1. Ear Science Institute Australia, Subiaco, Australia

2. Ear Sciences Centre, School of Medicine, The University of Western Australia, Nedlands, Australia

3. Department of Otolaryngology–Head and Neck Surgery, Sir Charles Gairdner Hospital, Perth, Australia

4. Department of Otolaryngology–Head and Neck Surgery, Stanford University, Stanford, California, USA

5. BRITElab, Harry Perkins Institute of Medical Research, QEII Medical Centre; Centre for Medical Research, The University of Western Australia, Nedlands, Australia

6. Department of Electrical, Electronic, and Computer Engineering, School of Engineering, The University of Western Australia, Nedlands, Australia

7. Institute for Frontier Materials, Deakin University, Geelong, Australia

8. Department of Speech Language Pathology and Audiology, University of Pretoria, Pretoria, South Africa

Abstract

Objective To evaluate the recent developments in optical coherence tomography (OCT) for tympanic membrane (TM) and middle ear (ME) imaging and to identify what further development is required for the technology to be integrated into common clinical use. Data Sources PubMed, Embase, Google Scholar, Scopus, and Web of Science. Review Methods A comprehensive literature search was performed for English language articles published from January 1966 to January 2018 with the keywords “tympanic membrane or middle ear,”“optical coherence tomography,” and “imaging.” Conclusion Conventional imaging techniques cannot adequately resolve the microscale features of TM and ME, sometimes necessitating diagnostic exploratory surgery in challenging otologic pathology. As a high-resolution noninvasive imaging technique, OCT offers promise as a diagnostic aid for otologic conditions, such as otitis media, cholesteatoma, and conductive hearing loss. Using OCT vibrometry to image the nanoscale vibrations of the TM and ME as they conduct acoustic waves may detect the location of ossicular chain dysfunction and differentiate between stapes fixation and incus-stapes discontinuity. The capacity of OCT to image depth and thickness at high resolution allows 3-dimensional volumetric reconstruction of the ME and has potential use for reconstructive tympanoplasty planning and the follow-up of ossicular prostheses. Implications for Practice To achieve common clinical use beyond these initial discoveries, future in vivo imaging devices must feature low-cost probe or endoscopic designs and faster imaging speeds and demonstrate superior diagnostic utility to computed tomography and magnetic resonance imaging. While such technology has been available for OCT, its translation requires focused development through a close collaboration between engineers and clinicians.

Publisher

SAGE Publications

Subject

Otorhinolaryngology,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3