Affiliation:
1. Vicenza Civil Hospital, Vicenza, Italy
Abstract
Objectives. To assess stability of a new-design titanium implant (Cochlear Baha BAI300) in bone-anchored hearing aid (Baha) recipients over time. Study Design and Setting. Prospective case series with planned data collection at a tertiary care institution. Twelve patients who underwent Baha surgery over 12 months were analyzed. Subjects. All patients (3 children, age range 6-10 years, mean 7.7 years, median 7.3 years; 9 adults, age range 34-73 years, mean 52.8 years, median 48 years) underwent the 1-stage procedure following the standard Brånemark technique. In all procedures, the single-skin incision was applied. Methods. The stability of the implant was measured with implant stability quotient (ISQ) measurement tests. All patients were tested at surgery and every week after for 1 year. Wound-healing time, degree of soft tissue reactions around the abutment, and need for revision surgery were examined. Two-tailed Student t test and χ2 for all comparisons were calculated; P < .05 values were considered significant. Results. The new implant showed a steep increase of ISQ values over the first 2 weeks and reached stable ISQ values 3 weeks after surgery. Complete skin healing time was obtained in 8 days on average (range, 7-12 days; median 7.6 days). Neither implant extrusion nor skin revision surgery was observed in our patients during the first year of follow-up. Conclusions. This new implant revealed stability and fast integration. The data from the present study suggest the new implant may be suitable for early loading protocols.
Subject
Otorhinolaryngology,Surgery
Cited by
31 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献