Comparison of Doppler ultrasonic sphygmomanometry, oscillometry and high-definition oscillometry for non-invasive blood pressure measurement in conscious cats

Author:

Knies Marieke12ORCID,Teske Erik2,Kooistra Hans2

Affiliation:

1. AniCura Veterinary Referral Centre Haaglanden, Dordrecht, The Netherlands

2. Department of Clinical Sciences, Faculty of Veterinary Medicine, Utrecht University, Utrecht, The Netherlands

Abstract

Objectives Systemic arterial hypertension is a common occurrence and can have serious adverse consequences in cats. Therefore, measuring blood pressure is very important. There are many indirect blood pressure measurement devices available. This study compared Doppler, oscillometric (petMAP Graphic II, SunTech Vet20, the Cardell Insight-X0000) and high-definition oscillometry devices for the non-invasive measurement of blood pressure in conscious cats. Methods In this prospective study, blood pressure was measured in 32 cats using the different devices according to the recommendations of the American College of Veterinary Internal Medicine Consensus Statement. Blood pressures (systolic blood pressure [SBP], diastolic blood pressure [DBP], mean arterial pressure [MAP]), time to complete measurements, number of attempts needed, coefficient of variation (CV) between the blood pressure measurements of the different devices, ease of completing measurements and apparent stress level of the cat were assessed. Results There was a significant difference between devices in the time taken to obtain blood pressure readings and the number of attempts necessary to obtain six reliable measurements. The CV of the Doppler device was significantly smaller than that of the rest of the devices, but there were no other differences between the devices. The mean SBP, DBP and MAP measured by the petMAP device were significantly higher than the measurements from the other devices. The perceived ease of measurement was not significantly different between the various machines. The perceived level of stress of measurement with the Doppler device was significantly higher compared with the other devices but did not lead to an increased SBP. Conclusions and relevance Using a Doppler device to measure blood pressure in conscious cats is fast, relatively easy and gives reliable results. A disadvantage is that the Doppler device can only measure SBP, while oscillometric devices also provide DBP and MAP. However, in veterinary medicine, systolic hypertension is considered the most relevant.

Funder

Boehringer Ingelheim Animal Health Netherlands B.V.

Publisher

SAGE Publications

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3