Reproducibility of Optic Nerve Head Measurements Obtained by Optical Coherence Tomography

Author:

Olmedo M.1,Cadarso-Suarez C.2,Gomez-Ulla F.13,Val C.1,Fernandez I.1

Affiliation:

1. Instituto Tecnológico de Oftalmología, Hospital Nuestra Señora de la Esperanza, Santiago de Compostela - Spain

2. Unit of Biostatistics, Department of Statistics and Operations Research Santiago de Compostela - Spain

3. Department of Ophthalmology, Complejo Hospitalario Universitario de Santiago, University of Santiago de Compostela, Santiago de Compostela - Spain

Abstract

Purpose To evaluate reproducibility of optic nerve head (ONH) morphometry measurements obtained by optical coherence tomography (the fast optical disk protocol of the Zeiss model 3000 OCT system) in normal and glaucomatous eyes. Methods Prospective instrument-evaluation study. ONH measurements for 20 eyes were obtained in eight scan sessions taken during two visits to an ophthalmology clinic (10 normal patients, 10 glaucoma patients, one eye per subject). At every one of the eight sessions for each eye, estimates of eight ONH morphometry variables (see Main outcome measures) were obtained. The first two sessions were performed by two operators, followed by a 30-minute break. The same operators then completed a third and fourth session. This sequence was duplicated on a second visit. Intrasession, intersession, intervisit, and interoperator reproducibility of the eight variables were calculated by the use of a components variance model. Intraclass correlation coefficients (ICC) were used to assess reliability. Results Vertical integrated rim area, horizontal integrated rim area, disk area, cup area, rim area, cup/disk area ratio, cup/disk horizontal ratio, cup/disk vertical ratio. With the exception of the horizontal integrated rim area and rim area in normal subjects, the factor subject was the most important source of variance for all variables. Reliability values as measured by ICC for normal eyes were above 81%, with the exception of measurements of the horizontal integrated rim area (23.1%), rim area (33.3%), and disk area (64.7%). For glaucomatous eyes all values were above 85%, with the exception of the disk area (68.1%). Conclusions ONH measurements obtained using the fast optical disk protocol of the Zeiss 3000 OCT system show good reproducibility, for both normal and glaucomatous eyes.

Publisher

SAGE Publications

Subject

Ophthalmology,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3