Icare versus Goldmann in a randomised middle-aged population

Author:

Stoor Katri12ORCID,Karvonen Elina12,Ohtonen Pasi23,Liinamaa M Johanna12,Saarela Ville12

Affiliation:

1. Department of Ophthalmology, Oulu University Hospital, Oulu, Finland

2. Medical Research Center Oulu, University of Oulu, Oulu, Finland

3. Division of Operative Care, Oulu University Hospital, Oulu, Finland

Abstract

Purpose The aim of this study was to compare the measurements of intraocular pressure by two tonometers, the Icare rebound tonometer and the Goldmann applanation tonometer, in a randomised screening study. The influence of refraction and central corneal thickness on the measurements was also evaluated. Methods Intraocular pressure was measured with rebound tonometer and Goldmann applanation tonometer in 1266 participants; refraction and central corneal thickness were also determined. One randomised eye of each participant was selected for this report’s analysis. A Bland–Altman plot was used to compare the values obtained with the two devices. Results The correlation between rebound tonometer and Goldmann applanation tonometer was good: the intraclass correlation coefficient (r) between the two methods was 0.735 ( p < 0.001). The mean difference (rebound tonometer–Goldmann applanation tonometer) was 0.11 ± 2.3 mmHg. The difference was not statistically significant (95% confidence interval: 0.11 to 0.13, p = 0.09). With increasing central corneal thickness, not only did intraocular pressure values with both devices increase, but the difference between them also increased. Refraction (spherical equivalent) did not influence intraocular pressure or the rebound tonometer–Goldmann applanation tonometer difference. However, high astigmatism (≥2D) exerted an influence on intraocular pressure values taken with Goldmann applanation tonometer. Conclusion Measurements with rebound tonometer and Goldmann applanation tonometer are relatively uniform although rebound tonometer slightly overestimated intraocular pressure. Both rebound tonometer and Goldmann applanation tonometer and the difference between these devices were affected by central corneal thickness but not by refraction. Higher astigmatism affected Goldmann applanation tonometer more than rebound tonometer. It is concluded that rebound tonometer is a reliable method for measuring intraocular pressure in a population-based screening study.

Publisher

SAGE Publications

Subject

Ophthalmology,General Medicine

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3