Agreement profiles for rebound and applanation tonometry in normal and glaucomatous children

Author:

Esmael Amanne1,Ismail Yomna M1,Elhusseiny Abdelrahman M1,Fayed Alaa E1,Elhilali Hala M1

Affiliation:

1. Department of Ophthalmology, Faculty of Medicine, Cairo University, Cairo, Egypt

Abstract

Objectives: To investigate agreement between intraocular pressure measurements by the rebound tonometer and handheld Perkins applanation tonometer in children with and without primary congenital glaucoma and test agreement with intraocular pressure and age variations. Materials and methods: A prospective non-interventional comparative study done on 223 eyes of 115 children, 161 normal eyes, and 62 eyes with primary congenital glaucoma. Intraocular pressure measurements were obtained in the upright position by rebound tonometer first, followed by installation of topical anesthetic eye drops (benoxinate), then measured by Perkins applanation tonometer. Results: For all eyes, mean difference between Perkins applanation tonometer and rebound tonometer was −0.59 ± 2.59 mmHg, p = 0.001. Regression analysis with (r) = 0.9, (r2) = 0.79, and p < 0.001. In primary congenital glaucoma: there was a mean difference of −.79 ± 2.82 (p = 0.032), a good correlation with (r) = 0.94, (r2) = 0.87%, and 95% level of agreement: –6.34 to +4.76. In normal eyes: mean difference was −.52 ± 2.5 (p = 0.01), correlation: (r) = 0.8, (r2) = 0.64, and p = 0.001. The 95% level of agreement −5.41 and +4.36 mmHg. In intraocular pressure ⩽ 15 mmHg: mean difference −0.89 ± 2.15 mmHg, 95% level of agreement between −5.1 and +3.32 mmHg, p < 0.001. In intraocular pressure >15 mmHg: mean difference was 0.04 ± 3.28 mmHg, 95% level of agreement −6.38 and +6.46 mmHg, p = 0.914. Conclusion: There is a good correlation between rebound tonometer and Perkins applanation tonometer in children with and without primary congenital glaucoma; however, rebound tonometer overestimates the intraocular pressure, and in intraocular pressure >15 mmHg there is less agreement between the two devices. Hence, in higher intraocular pressure measurement caution should be taken when interpreting rebound tonometer readings, and a confirmatory measurement using Perkins applanation tonometer is advised.

Publisher

SAGE Publications

Subject

Ophthalmology,General Medicine

Cited by 22 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3