Comparison of central corneal thickness and intraocular pressure measured with two different tono/pachymeter devices in non-glaucomatouse children

Author:

Teberik Kuddusi1ORCID,Eski Mehmet Tahir2,Ankarali Handan3

Affiliation:

1. Department of Ophthalmology, Düzce University Medical School, Düzce, Turkey

2. Private Neon Hospital, Erzincan, Turkey

3. Biostatistics and Medical Informatics Department, İstanbul Medeniyet University Faculty of Medicine, Istanbul, Turkey

Abstract

Purpose: The goal of this research is to compare the intraocular pressure (IOP) and the mean central corneal thickness (CCT) values obtained from the measurements with Nidek NT-530P and Canon TX-20P devices with the values obtained by ultrasound pachymetry (UP) and Goldmann applanation tonometry (GAT) in children. Methods: This prospective study was conducted with 119 healthy children. The measurements were repeated three times for each eye. The intraclass correlation coefficient (ICC) was used to assess the correlation between the measurements obtained from different devices. The Bland–Altman plot was used to analyze the agreement between two different devices graphically. The measurements were taken in the same order in all subjects: TX-20P, NT-530P, UP, and GAT. Results: The mean age of the children was 10.1 ± 3.2 (6–17) years. The mean CCT values for the eyes were 568.90 and 569.68 (TX-20P), 571.44 and 566.37 (NT-530P), 564.77 and 564.67 μm UP, (right and left, respectively). The highest correspondence observed for CCT was between UP and NT-530P devices (ICC, 0.982, the left eye). The mean IOP measurements for Canon TX-20P, NT-530P, and GAT were 16.5 ± 3.2, 16.3 ± 3.2, and 16.8 ± 3.6 mmHg for the right eyes, respectively, while those for the left eyes were 16.6 ± 3.1, 16.2 ± 3.3, and 16.8 ± 3.7 mmHg. The highest correspondence was noted between measurements obtained using GAT and Nidek NT-530P devices (ICC, 0.945, in the left eye). Conclusion: Tono/pachymetry devices in the healthy children are provides significant agreements with the gold standard methods for IOP and CCT measurements. Clinical Trial Registration number: 2016-94

Publisher

SAGE Publications

Subject

Ophthalmology,General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Pediatric glaucoma in aniridia in a population of Egyptian children;Journal of American Association for Pediatric Ophthalmology and Strabismus;2023-08

2. Thin Central Corneal Thickness May Be a Risk Factor for Myopia Progression in Children;Journal of Ophthalmology;2023-01-16

3. Comparison of central corneal thickness measurements with different non-contact devices in healthy children;Photodiagnosis and Photodynamic Therapy;2022-12