Affiliation:
1. G.B. Bietti Eye Foundation, Roma - Italy
Abstract
A prospective study was conducted on 90 eyes undergoing extracapsular cataract extraction and posterior chamber IOL implantation. Postoperative IOL decentration was evaluated with respect to the loop material, either polypropylene (prolene) or polymethylmethacrylate (PMMA), and the IOL fixation technique, either in the capsular sac (in-the-bag) after capsulorhexis, or in the ciliary sulcus after “can-opener” capsulotomy. More decentration was observed with IOL having polypropylene loops than with those having PMMA loops (0.80 ± 0.735 mm versus 0.373 ± 0.299 mm). IOL implanted in the bag showed greater decentration than those implanted in the sulcus, but the difference was not significant. Our findings confirm the recommendation to use an IOL with PMMA loops when placing the implant in the capsular sac.
Subject
Ophthalmology,General Medicine
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献