Affiliation:
1. Department of Ophthalmology, AFMC, Pune, Maharashtra, India
Abstract
Objective To report a case of post radial keratotomy (RK) cataract in a 55-year-old lady wherein biometry was done by ray-tracing method incorporated in scheimpflug topographer (Sirius + Scheimpflug Analyzer, CSO, Italy). Method In our case, we performed intraocular lens (IOL) power calculation using a recent concept of ray tracing with scheimpflug topographer and compared with traditional methods available at American Society of Cataract and Refractive Surgery(ASCRS) website ( www.ascrs.org ) for eyes with prior RK. Phacoemulsification was performed and a monofocal + 24.5D IOL implanted in the capsular bag. Result Manifest refraction at six weeks postoperative period was + 1.0DS/-2.0DC × 75° with spherical equivalence of 0. On comparison of all the methods used to calculate IOL power, the absolute errors of ray tracing and Barrett true K were found to be the least, 0.14 and 0.18 respectively. Conclusion Ray tracing biometry with scheimpflug topographer seems to provide accurate IOL power in post RK eyes.
Subject
Ophthalmology,General Medicine