Affiliation:
1. Faculty of Optics and Optometry, Complutense University of Madrid, Spain
2. Fernández-Vega Ophthalmological Institute, Oviedo, Spain
Abstract
Purpose: To objectively analyze the optical quality of the FineVision Toric intraocular lens (IOL) with two cylinder powers when different combinations of rotations and residual refractive errors are induced. Methods: This study assessed the FineVision Toric IOL with two different cylinder powers: 1.5 and 3.0 diopters (D). Three different rotation positions were considered: centered, 5° and 10° rotated. An optical bench (PMTF) was used for optical analysis. The optical quality of the IOLs was calculated by the modulation transfer function (MTF) at five different focal points (0.0, 0.25, 0.50, 0.75, and 1.00 D). Results: The MTF averaged value of the reference situation was 38.58 and 37.74 for 1.5 and 3.0 D of cylinder, respectively. For the 1.5 D cylinder, the combination of 5° of rotation with a defocus of 0.25, 0.50, 0.75, and 1.0 D induced a decrease on the MTF of 12.39, 19.94, 23.43, 24.23 units, respectively. When induced rotation was 10°, the MTF decrease was 17.26, 23.40, 24.33, 24.48 units, respectively. For the 3.0 D cylinder, the combination of 5° with 0.25, 0.50, 0.75, and 1.0 D of defocus, induced a decrease on the MTF of 12.51, 18.97, 22.36, 22.48 units, respectively. When induced rotation was 10°, the MTF decrease was: 18.42, 21.57, 23.08, and 23.61 units, respectively. Conclusion: For both FineVision Toric IOLs there is a certain optical tolerance to rotations up to 5° or residual refractive errors up to 0.25 D. Situations over these limits and their combination would affect the visual quality of patients implanted with these trifocal toric IOLs.
Subject
Ophthalmology,General Medicine
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献