Comparison of Icare HOME and non-contact tonometer in intraocular pressure measurement in the early stage after ICL V4c implantation

Author:

Ye Yuhao123,Yang Fangwen4,Ding Lan123,Xia Fei123,Niu Lingling123,Wang Xiaoying123,Zhao Jing123,Zhou Xingtao123ORCID

Affiliation:

1. Department of Ophthalmology and Optometry, Eye & ENT Hospital, Fudan University, Shanghai, China

2. NHC Key Laboratory of Myopia (Fudan University), Key Laboratory of Myopia, Chinese Academy of Medical Sciences, Shanghai, China

3. Shanghai Research Center of Ophthalmology and Optometry, Shanghai, China

4. The second affiliated Hospital of Guizhou Medical University, Guizhou, China

Abstract

Purpose To investigate the characteristics of intraocular pressure (IOP) measurements using the Icare HOME rebound tonometer (RBT) and non-contact tonometer (NCT) during the early stage after implantable collamer lens (ICL) V4c implantation, and to assess the agreement between the two methods. Methods This prospective case series study included 104 eyes of 53 patients (mean age 28.77  ±  5.34 years), who underwent ICL V4c implantation. IOP was measured preoperatively, and at 0.5, 1, 2, 4, and 24 h postoperatively by Icare HOME and NCT, respectively. Results All surgeries were uneventful. IOP measurements by Icare and NCT preoperatively were 14.03  ±  2.90 mmHg and 14.09  ±  2.87 mmHg, respectively.The corresponding values were 12.56  ±  8.09 mmHg and 14.12  ±  6.52 mmHg (P > 0.05) at postoperative 0.5 h, 19.21  ±  8.74 mmHg and 19.60  ±  7.66 mmHg at postoperative 1 h (P > 0.05), 21.21  ±  8.10 mmHg and 20.31  ±  6.93 mmHg at postoperative 2 h (P > 0.05), 16.11  ±  5.89 mmHg and 17.04  ±  4.84 mmHg at postoperative 4 h (P > 0.05), and 14.04  ±  3.88 mmHg and 14.78  ±  2.80 mmHg at postoperative 24 h, respectively (P > 0.05). There was good agreement based on intraclass correlation coefficients (ICCs) between NCT and Icare HOME (all ICCs > 0.6 at different time points [range, 0.6986–0.956]). The Bland-Altman plot showed a mean percentage of over 95.81% of the points falling within the limits of agreement. There was a significant difference in the low IOP measurements (<10 mmHg) between Icare HOME and NCT (7.28  ±  2.55 vs. 10.65  ±  2.48, P < 0.001). Conclusion Icare HOME can be used for IOP measurement after ICL V4c implantation and demonstrated good agreement with NCT, except in cases with low intraocular pressure (<10 mmHg).

Funder

Project of Shanghai Xuhui District Science and Technology

Project of Shanghai Science and Technology

Clinical Research Plan of SHDC

National Natural Science Foundation of China

Joint research project of new frontier technology in municipal hospitals

Publisher

SAGE Publications

Subject

Ophthalmology,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Refractive surgery for the glaucoma suspect;Current Opinion in Ophthalmology;2023-03-29

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3