Perineural Entrapment of an Interscalene Stimulating Catheter

Author:

Adhikary S. D.1,Armstrong K.1,Chin K. J.1

Affiliation:

1. Department of Anesthesiology and Perioperative Medicine, University of Western Ontario, London, Ontario, Canada

Abstract

A patient undergoing elective total shoulder arthroplasty had an Arrow Stimucath™ (Arrow International Reading, PA, USA) stimulating catheter inserted in the region of the interscalene brachial plexus using an ultrasound-guided in-plane approach. There was subsequent difficulty in removing the catheter and traction was associated with painful paraesthesia in the patient's thumb. Plain X-ray revealed a hook-shaped deformity of the tip and surgical exploration was required to remove the catheter, the tip of which appeared to be trapped within the sheath of the brachial plexus. We speculate that the mechanism for entrapment in this case was deformation of the catheter tip into a hook-like shape. The subsequent catheter-neural interaction prevented asymptomatic removal. We recommend that removal of peripheral nerve catheters be attempted only after resolution of sensory block so as to enable patients to report pain or paraesthesia. Imaging with ultrasound or X-rays may help ascertain catheter tip location and confirm whether deformity is present. We also recommend a low threshold for proceeding to surgical extraction, particularly if neurological symptoms are present.

Publisher

SAGE Publications

Subject

Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3