An Evaluation of Five Carbon Dioxide Analysers for use in the Operating Theatre and Intensive Care Unit

Author:

Selby D. G.1,Ilsley A. H.12,Runciman W. B.13

Affiliation:

1. Department of Anaesthesia and Intensive Care, The Flinders University of South Australia, Flinders Medical Centre, South Australia

2. Principal Hospital Scientist/Lecturer.

3. Senior Staff Specialist/Senior Lecturer.

Abstract

Five commercially available carbon dioxide analysers were assessed with respect to accuracy, response time, stability, the effect of water vapour and the effect of rebreathing. Two side-stream sampling analysers (Datex Normocap and Engstrom Eliza) and one main-stream sampling analyser (Hewlett-Packard Capnoshot) were found to be satisfactory for use both in the operating theatre and intensive care unit. The Instrumentation Laboratory (IL) 200, a side-stream monitor, performed satisfactorily but neither it nor the Siemens Sirecust 404, a main-stream monitor, has the facility to compensate for the presence of nitrous oxide. The Siemens analyser, in the form tested, could not be recommended for clinical use, as it was less accurate than the other monitors, could not detect rebreathing, and was subject to fluctuations when used with gas mixtures saturated with water vapour.

Publisher

SAGE Publications

Subject

Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3