Affiliation:
1. Department of Anesthesiology, Shanghai Eye, Ear, Nose, and Throat Hospital, Fudan University, Shanghai, China
Abstract
This prospective, randomised study was conducted to assess the effect of flexible laryngeal mask airway (FLMA) size on oropharyngeal leak pressure (OLP) in children at the recommended intracuff pressure. A total of 120 children undergoing elective ophthalmic surgery were randomly assigned to the size 2 FLMA group or size 2.5 FLMA group. The primary measurement was OLP at an intracuff pressure of 40 cmH2O. Secondary outcomes included the incidence of OLP <10 cmH2O, insufficient ventilation, gastric insufflation, insertion time, successful first-attempt insertion rate, fibreoptic view grade and pharyngolaryngeal adverse events. The median OLP was comparable for the size 2 and size 2.5 FLMA (18 cmH2O versus 18 cmH2O, P=0.38). However, the size 2 FLMA group had a higher incidence of OLP <10 cmH2O and insufficient ventilation (13.3% versus 0, P=0.006). In subgroup analyses based on weight, the size 2.5 FLMA had a lower occurrence of OLP <10 cmH2O and insufficient ventilation (27% versus 0, P=0.0046) in children 16–20 kg. We conclude that at a 40 cmH2O intracuff pressure, the OLP with the size 2 and size 2.5 FLMA was similar in children weighing 10–15.9 kg. However, in children weighing 16–20 kg, size 2 devices had a higher incidence of low OLP and insufficient ventilation, so a 2.5 FLMA may be preferable in this subgroup.
Subject
Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献