Telerobotic ultrasound to provide obstetrical ultrasound services remotely during the COVID-19 pandemic

Author:

Adams Scott J1ORCID,Burbridge Brent1ORCID,Chatterson Leslie1,McKinney Veronica2,Babyn Paul1,Mendez Ivar3

Affiliation:

1. Department of Medical Imaging, University of Saskatchewan, Canada

2. Northern Medical Services, Department of Academic Family Medicine, University of Saskatchewan, Canada

3. Department of Surgery, University of Saskatchewan, Canada

Abstract

Introduction Obstetrical ultrasound imaging is critical in identifying at-risk pregnancies and informing clinical management. The coronavirus disease 2019 (COVID-19) pandemic has exacerbated challenges in accessing obstetrical ultrasound for patients in underserved rural and remote communities where this service is not available. This prospective descriptive study describes our experience of providing obstetrical ultrasound services remotely using a telerobotic ultrasound system in a northern Canadian community isolated due to a COVID-19 outbreak. Methods A telerobotic ultrasound system was used to perform obstetrical ultrasound exams remotely in La Loche, Canada, a remote community without regular access to obstetrical ultrasound. Using a telerobotic ultrasound system, a sonographer 605 km away remotely controlled an ultrasound probe and ultrasound settings. Twenty-one exams were performed in a five-week period during a COVID-19 outbreak in the community, including limited first-, second- and third-trimester exams ( n = 11) and complete second-trimester exams ( n = 10). Participants were invited to complete a survey at the end of the telerobotic ultrasound exam describing their experiences with telerobotic ultrasound. Radiologists subsequently interpreted all exams and determined the adequacy of the images for diagnosis. Results Of 11 limited obstetrical exams, radiologists indicated images were adequate in nine (81%) cases, adequate with some reservations in one (9%) case and inadequate in one (9%) case. Of 10 second-trimester complete obstetrical exams, radiologists indicated images were adequate in two (20%) cases, adequate with some reservations in three (30%) cases and inadequate in five (50%) cases. Second-trimester complete obstetrical exams were limited due to a combination of body habitus, foetal lie and telerobotic technology. Discussion A telerobotic ultrasound system may be used to answer focused clinical questions such as foetal viability, dating and foetal presentation in a timely manner while minimising patient travel to larger centres and potential exposure to severe acute respiratory virus coronavirus 2 during the COVID-19 pandemic.

Funder

Saskatchewan Health Research Foundation / Saskatchewan Centre for Patient-Oriented Research

Publisher

SAGE Publications

Subject

Health Informatics

Cited by 26 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献