Comparative preclinical evaluation of a polymer-free sirolimus-eluting stent in porcine coronary arteries

Author:

Sperling Christian1,Waliszewski Matthias W.12ORCID,Kherad Behrouz2,Krackhardt Florian3

Affiliation:

1. Medical Scientific Affairs, B. Braun Melsungen AG, Berlin, Germany

2. Department of Internal Medicine and Cardiology, Charité – Universitätsmedizin Berlin, Berlin, Germany

3. Department of Cardiology, Charité – Universitätsmedizin Berlin, Campus Virchow, Augustenburger Platz 1, D-13353 Berlin, Germany

Abstract

Background: Polymer-free drug-eluting stents (DES) without permanent-polymer coating may be associated with rapid vessel healing, providing a rationale to reduce dual-antiplatelet therapy (DAPT). The aim of the current study was to compare vessel healing of a polymer-free sirolimus-eluting stent (PF-SES), its bare metal stent (BMS) analogue to a permanent polymer-based sirolimus-eluting stent (SES) with proven effectiveness in porcine coronary arteries. Material and methods: An ultrathin-strut cobalt–chromium PF-SES, its BMS analogue and an SES with a permanent polymer were used to study vessel healing and their antistenotic potential. Stents were implanted in porcine coronary arteries for histopathologic analysis at 7, 28 and 180 days. In an additional in vitro study, the thrombogenicity of PF-SES was compared with a fluoropolymer-coated everolimus-eluting stent (EES) which demonstrated low stent thrombosis rates in numerous studies. Results: In the animal study, neointimal growth and injury scores were minimal and inflammation scores were low in the neointima and adventitia in all study groups. After 28 days, neointimal area was lowest in PF-SES when compared with SES and BMS (1.48 ± 0.55 mm² versus 2.43 ± 0.69 mm² versus 1.90 ± 0.85 mm², respectively, p < 0.05) and endothelialization of luminal surfaces was nearly complete in all groups, though SES show the least coverage with occasional adherent luminal inflammatory cells ( p > 0.05). At 180 days, neointimal area and thickness were most pronounced in SES ( p < 0.05) and comparable with BMS implantations, which were characterized by nearly completed vessel healing. PF-SES and BMS had complete endothelialization, absence of fibrin and sustained low inflammatory reaction when compared with the permanent polymer-based SES (inflammation score: PF-SES 0.41 ± 0.74 versus SES 2.52 ± 1.72 versus BMS 0.30 ± 0.65, respectively, p < 0.05 BMS versus SES). Granuloma formation and fibrin accumulation were most pronounced in SES but did not reach statistical significance, p > 0.05). In the in vitro thrombogenicity study, the PF-SES confirmed comparable antithrombogenic properties with regard to the parameters fibrin and platelet binding, and platelet aggregation when compared with the EES. Conclusions: As compared with BMS, the ultrathin-strut cobalt–chromium PF-SES showed similar endothelialization at 28 days and comparable healing characteristics at 180 days efficacious inhibition of neointimal proliferation in porcine coronary arteries with low inflammation responses and a BMS-like endothelialization at 180 days. In addition, in an in vitro model, the PF-SES also confirmed low thrombogenicity as compared with the EES.

Publisher

SAGE Publications

Subject

Pharmacology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3