Affiliation:
1. CEDIMAT, Plaza de la Salud; Santo Domingo, Republica Dominicana
2. Philips Healthcare; Tamboré, Barueri, Brazil
Abstract
The assessment of shunt reduction after an embolization of an arteriovenous malformation (AVM) or fistula (AVF) from conventional angiography is often difficult and may be subjective. Here we present a completely non-invasive method using magnetic resonance imaging (MRI) to measure shunt reduction. Using pulsed arterial spin labeling (PASL), we determined the relative amount of signal attributed to the shunt over 1.75 s and 6 different slices covering the lesion. This amount of signal from the shunt was related to the total signal from all slices and measured before and after embolization. The method showed a fair agreement between the PASL results and the judgement from conventional angiography. In the case of a total or subtotal shunt occlusion, PASL showed a shunt reduction between 69% and 92%, whereas in minimal shunt reduction as judged by conventional angiography, the ASL result was −6% (indicating slightly increased flow) to 35% in a partially occluded vein of Galen aneurysm. The PASL method proved to be fairly reproducible (up to 2% deviation between three measurements without interventions). On conclusion, PASL is able to reliably measure the amount of shunt reduction achieved by embolization of AVMs and AVFs.
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献