Abstract
The use of flow diverting stents for wide based, intracranial aneurysms has become an invaluable treatment option. While intracranial hemorrhage and ischemic stroke from dislodged atherosclerotic emboli are common adverse events, the potential for delayed granulomatous inflammation from possible hydrophilic polymer emboli is rarely recognized. We present a unique case in which visible chipping of the pusher wire for stent placement was observed, followed by clinical and radiographic evidence suggestive of a delayed foreign body reaction to intracranial hydrophilic polymer emboli. A 55-year-old woman underwent placement of a Pipeline embolization device for a left-sided, broad-based aneurysm at the base of the internal carotid artery and posterior communicating artery. Two months later she developed right-sided focal neurological deficits. Imaging showed ipsilateral focal edema and enhancing lesions with contrast. Although not confirmed with biopsy and histopathology, clinical and radiographic evidence suggests that this patient probably experienced a delayed foreign body reaction to hydrophilic polymer emboli from compromised procedural equipment during flow diverting stent placement. Although previously described, this is the first instance to our knowledge in whichvisible chipping of the pusher wire was observed on a Pipeline embolization device.