Evaluation of Operator Radioprotection Using a New Injection Device during Vertebroplasty

Author:

Nguyen-Kim L.1,Fargeot C.2,Beaussier H.3,Payen S.4,Chiras J.5

Affiliation:

1. Pharmacy Department, Medical Device Evaluation and Purchasing Department; AGEPS, AP-HP; Paris, France

2. Pharmacy Department, Pitie Salpetrière Hospital, AP-HP; Paris, France

3. Pharmacy Department; Saint-Joseph Hospital, AP-HP; Paris, France

4. Radioprotection Department, Pitie Salpetrière Hospital, AP-HP; Paris, France

5. Neuroradiology Department, Pitie Salpetrière Hospital, AP-HP; Paris, France

Abstract

This study aimed to evaluate the protection granted by a simple device (X'TENS®, Thiebaud, France) and to provide operators with information on the performance of this new device, which has not yet been assessed. Our assumption is that this device efficiently reduces the radiation dose to the operator. In a prospective clinical study, the radiation dose the operator's hand receives has been assessed using a specific sensor (UNFOR Instrument). Each patient included in the study was to receive at least two injections of cement during the procedure. Exposure was measured with and without the range extender. The data collected were then processed using a Wilcoxon matched pairs test. During 14 interventions, 20 vertebrae were treated with both procedures. Eleven women and three men were included. Seven patients underwent vertebroplasty for metastatic lesions and seven for osteoporotic lesions, bone fractures or vertebral compressions. The average injection time was 1.35 minutes with the device and 1.20 without (p=0.75). The dose to the hand per ml injected was 111.37 vs. 166.91 (p<0.05). Theoretically, the protection granted by the range extender depends on the length of the device. Our results are consistent with the inverse-square law. However, the variations in our results indicate that a proper and rigorous use is mandatory for the device to be effective. Given that radioprotection during fluoroscopy procedures is a frequently raised issue, the need for information for a safer practice increases likewise.

Publisher

SAGE Publications

Subject

Immunology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3