Staff dose evaluation by application of radiation protection during endoscopic retrograde cholangiopancreatography (ERCP) procedures performed with a mobile C-arm

Author:

Österlund Andreas1ORCID,Drohn Wilfried2,Hoedlmoser Herbert3ORCID,Greiter Matthias3,Schmid Martin4,Källman Hans-Erik1

Affiliation:

1. Imaging and Functional Medicine, Region Dalarna, Falun, Sweden

2. Department of Internal Medicine, Falu Hospital, Falun, Sweden

3. Helmholtz Zentrum München, Individual Monitoring Service, Munich, Germany

4. MAVIG GmbH, Munich, Germany

Abstract

Background The radiation dose to staff performing endoscopic retrograde cholangiopancreatography (ERCP) is not negligible. Purpose To evaluate the shielding effect of a table-suspended lower-body radiation shield for the positions in the room occupied by the operator, assisting nurse, and anesthesiologist, used during ERCP procedures with a mobile C-arm. Material and Methods Eye lens dose, whole body dose, and extremity dose were measured with and without a table-suspended lower-body radiation shield in a phantom model and in clinical routine work. The effect of the shield was evaluated for each scenario and compared, and a projection was made for when shielding should be required from a regulatory point of view. Results In the phantom measurements, the shield provided significant shielding effects on the body and lower extremities for the operator but no significant shielding of the eye lens. The shielding effect for the assisting nurse was limited to the lower extremity. The clinical measurements yielded the same general result as the phantom measurements, with the major difference that the shield provided no significant reduction in the whole-body dose to the operator. Conclusion The table-suspended shield has a significant shielding effect for the lower extremities of the operator and assisting nurse. For annual dose–area product values >300,000 cGycm2, the protection of the operator should be reinforced with a ceiling-suspended shield to avoid doses to the eye lens and body in excess of regulatory dose restrictions.

Publisher

SAGE Publications

Subject

Radiology, Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

Reference26 articles.

1. Effectiveness of additional lead shielding to protect staff from scattering radiation during endoscopic retrograde cholangiopancreatography procedures

2. 3. ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 2007;37:1–332.

3. European Parliament. Council Directive 2013/59/Euratom of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation, and repealing Directives 89/618/Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/Euratom and 2003/122/Euratom. Available at: https://eur-lex.europa.eu/eli/dir/2013/59/oj.

4. Swedish Radiation Safety Authority. Strålsäkerhetsmyndighetens författningssamling 2018:1. Available at: https://www.stralsakerhetsmyndigheten.se/contentassets/edd48d6fa0114e9cb3ae07f3956babcc/ssmfs-20181-stralsakerhetsmyndighetens-foreskrifter-om-grundlaggande-bestammelser-for-tillstandspliktig-verksamhet-med-joniserande-stralning-konsoliderad-version.pdf

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献