Affiliation:
1. Department of Radiology, County Hospital, S-301 85 Halmstad, Sweden
Abstract
Purpose: The aim of the study was to investigate the performance of different types of equipment for bone mineral assessment in Sweden and Norway with standardised phantoms. Material and Methods: Measurements of the bone mineral content of 2 semi-anthropomorphic phantoms, one simulating the human torso and one the lower arm, at 21 hospitals in Norway and Sweden, were undertaken. The phantoms had inserts of hydroxyapatite in different concentrations and were measured together with additions of stearic acid simulating fat. The phantoms were repeatedly measured on 14 dual X-ray absorptiometry, 5 single X-ray absorptiometry, and 3 single photon absorptiometry devices and 1 quantitative CT device. Results: The measurements showed a large spread in the measured hydroxyapatite concentrations among different devices. The inaccuracy was slightly lower than 10% without fat present when the results were expressed as bone mineral areal mass (BMA, g/cm2). With fat included the inaccuracy was at worst 50%. Conclusion: The large inaccuracy found in this study gives rise to a large overlap between healthy and osteoporotic cases and makes the diagnosis of osteoporosis difficult in the individual patient.
Subject
Radiology Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology
Cited by
1 articles.