Author:
Yang X. M.,Manninen H.,Ji H. X.,Vainio P.,Soimakallio S.
Abstract
To evaluate the accuracy of digital videodensitometric technique in directly quantitating concentration of contrast medium, iohexol 300 mg I/ml was injected into a 2-mm-diameter plastic tube, in which clean water was circulated at a 190 ml/min flow, for digital subtraction angiography. Altogether 27 injections were performed with 3, 4 and 5 ml volumes at 3-, 4- and 5-ml/s flows of the contrast medium. A time-density curve was achieved by selecting a “vessel” region of interest (ROI) and a background ROI. Then, a frame corresponding to the maximum opacification of the contrast medium could be calculated. Finally, the average density and the time to peak density of the contrast medium were obtained. The average density was statistically higher (p < 0.01) with 5 ml/s flow than with 4- and 3-ml/s flows. Times to peak density reduced as injection flows or volumes increased. The results support the conclusion that digital videodensitometric technique is an accurate method for quantitation of contrast medium concentration during angiography. The angiographic opacification may be improved by injecting the iodine contrast medium with higher flows or larger volumes.
Subject
Radiology, Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology