Long-term integrity of 53 iliac vein stents after catheter-directed thrombolysis

Author:

Foegh Pia1ORCID,Strandberg Charlotte2,Joergensen Sanne2,Myschetzky Peter S2,Klitfod Lotte1,Just Sven2,Hansen Susanne3,Baekgaard Niels1

Affiliation:

1. Department of Vascular Surgery, Gentofte Hospital and Rigshospitalet, University of Copenhagen, Denmark

2. Department of Diagnostic Radiology, Herlev and Gentofte Hospital, University of Copenhagen, Denmark

3. Center for Clinical Research and Prevention, Bispebjerg and Frederiksberg Hospital, University of Copenhagen, Denmark

Abstract

Background Long-term surveillance data on venous stent integrity is sparse. There is limited knowledge on whether duplex ultrasound (DUS) can detect potential stent deformities such as kinking, straightening, and fracture, which may impact long-term patency of the stented veins. Purpose To assess venous stent integrity after at least five years of follow-up and to establish the efficacy of DUS as surveillance in patients with venous stent. Material and Methods A total of 45 patients with acute iliac-femoral deep vein thrombosis (DVT) treated with catheter directed thrombolysis (CDT) and stenting >5 years before follow-up. Stents were evaluated with 3D volume low dose non-contrast computed tomography (CT) and DUS for kinking, straightening, stent fracture, and patency. Results from CT scans and DUS were compared to assess the overall agreement between the methods. Results Median follow-up was 13.2 years (mean = 11.2 years; range = 5.2–15.8 years). 3D CT reconstructions showed normal stent configuration in 47 stents (89%). All intact stents were identified by DUS. In the remaining six stents, 3D CT reconstructions showed compression, tapering, kinking, and minor fracture. DUS recognized all stent complications except the minor fracture. Overall agreement between CT and DUS was 98% (kappa = 0.90). Two cases of stent occlusion were found. Conclusion The long-term physical resilience of iliac vein stents evaluated with 3D CT in patients treated with CDT for iliofemoral DVT was high. Stent deformities were mostly compression, whereas fracture was rarely seen. DUS seems to be sufficient to evaluate venous stent integrity.

Publisher

SAGE Publications

Subject

Radiology, Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3