Therapeutic angiographic procedures: differences in dose area product between analog image intensifier and digital flat panel detector

Author:

Spira Daniel1,Kirchner Sebastian2,Blumenstock Gunnar3,Herz Klaus4,Ketelsen Dominik2,Wiskirchen Jakub5,Wiesinger Benjamin2

Affiliation:

1. Department of Diagnostic and Interventional Radiology, Universitätsklinik Heidelberg, Heidelberg, Germany

2. Department of Diagnostic and Interventional Radiology, Eberhard-Karls-University, Tübingen, Germany

3. Institute of Clinical Epidemiology and Applied Biometry, Eberhard-Karls-University, Tübingen, Germany

4. Institute for Radiation Protection and Radionuclide Laboratory, Eberhard-Karls-University, Tübingen, Germany

5. Department of Radiology and Nuclear Medicine, Franziskus Hospital, Bielefeld, Germany

Abstract

Background Radiation exposure remains an unceasing concern in angiographic procedures. Modern angiography machines such as analog image intensifiers (AII) or the new flat panel detectors (FPD) aim at a further dose reduction. Purpose To present dose area products (DAP) in a broad spectrum of therapeutic angiographic procedures, comparing an AII to an FPD angiography system. Material and Methods A total of 999 peripheral therapeutic angiography procedures performed with an FPD ( n = 562) and an AII system ( n = 437) were evaluated. DAP, fluoroscopy time, and patients' body mass index (BMI) were recorded. Interventions were classified into five main groups: percutaneous transluminal angioplasty (PTA); PTA and stent placement; intra-arterial thrombolysis; embolization procedures; and specialized interventions. Results DAP values in therapeutic angiographic procedures were significantly higher when performed with the FPD compared to the AII system. The increase of the FPD versus AII system was 100.1% for PTA, 39.9% for PTA and stent placement, 187% for intra-arterial thrombolysis, 31.3% for embolization procedures, and 361% for specialized interventions. These differences persisted after standardizing DAP values to the geometric mean fluoroscopy duration of each procedure. Fluoroscopy times were shorter in all interventions performed at the FPD as compared to the AII system. DAPs increased with higher BMI, but the DAP increase of both systems with elevated BMI was variable, depending on the individual intervention. Conclusion In therapeutic angiographic procedures, the FPD system required higher DAPs despite shorter fluoroscopy times as compared to an AII system. Better ergonomics and speediness of the FPD system may be advantageous in the emergency setting.

Publisher

SAGE Publications

Subject

Radiology Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

Cited by 13 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3