Abstract
Lithium fluoride (LiF) thermoluminescent dosemeters (TLD) have been employed to measure the radiation dose distribution within a phantom and the central axis dose in air. Results are presented for seven (four EMI CT1010, one EMI CT5005, one EMI CT7070 and one Siemens DR2) different machines. Organ doses for four different computed tomography (CT) investigations (head, lung, liver and pelvis) have been estimated from previously published tables and the central axis dose in air in 3 CT units. These estimated organ doses were used in turn to deduce the somatic and genetic risks for the four types of CT examination. These calculations indicate that for an ‘average’ male patient undergoing a CT examination corresponding to the average conditions encountered, the total somatic risks are 3.15 10−4, 1.98 10−4, 2.31 10−4, and 1.38 10−4 for head, lung, liver and pelvis scans, respectively. The corresponding figures for female patients are 3.39 10−4, 3.95 10−4, 2.73 10−4, and 1.60 10−4. The risk from head scanning is approximately 250 times that of a dental pantomograph. Somatic and genetic risks will be approximately twice as high for contrast examinations.
Subject
Radiology, Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献