Pressure and stiffness of a new air-inflated compression wrap

Author:

Mosti Giovanni1ORCID,Gasparis Antonios2,Oropallo Alisha2,Labropoulos Nicos3ORCID

Affiliation:

1. Angiology Department, Clinica MD Barbantini, Lucca, Italy

2. Department of Surgery, Northwell Health, New Hyde Park, NY, USA

3. Division of Vascular Surgery, Stony Brook Medicine, Stony Brook, NY, USA

Abstract

Objective To report pressure and stiffness, in healthy volunteers, of a new compression device with an air bladder inflated by a pump to regulate pressure. Methods The device was applied to 60 legs of 30 volunteers and set to exert different pressures of 20–50 mmHg. The exerted pressure was measured in supine and standing positions and during simple physical exercises; static stiffness index, dynamic stiffness index, and walking pressure amplitudes were calculated. Results The exerted pressure showed a good correlation with the expected pressure at each pressure range. The stiffness indices were >10 mmHg in the range of inelastic materials. The device was considered very easy to apply and use by the testing researchers. Conclusions The device stiffness is in the same range as the inelastic bandages. Consequently, similar hemodynamic effectiveness could be expected but must be proved. Unlike inelastic bandages, this device was easy to apply and use.

Publisher

SAGE Publications

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3