Development of Automated Delivery Quality Assurance Analysis Software for Helical Tomotherapy

Author:

Yoon YH12ORCID,Shin Han-Back1,Han MC1,Kim Hojin1,Kim DW1,Hong Chae-Seon1ORCID,Kim Jihun3,Kim JS124

Affiliation:

1. Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, South Korea

2. Medical Physics and Biomedical Engineering Lab (MPBEL), Yonsei University College of Medicine, Seoul, South Korea

3. Department of Radiation Oncology, Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul, Korea

4. Oncosoft Inc., Seoul, South Korea

Abstract

BackgroundTo develop a fully automated in-house gamma analysis software for the “Cheese” phantom-based delivery quality assurance (QA) of helical tomotherapy plans.MethodsThe developed in-house software was designed to automate several procedures, which need to be manually performed using commercial software packages. The region of interest for the analysis was automatically selected by cropping out film edges and thresholding dose values (>10% of the maximum dose). The film-measured dose was automatically aligned to the computed dose using an image registration algorithm. An optimal film scaling factor was determined to maximize the percentage of pixels passing gamma (gamma passing rate) between the measured and computed doses (3%/3 mm criteria). This gamma analysis was repeated by introducing setup uncertainties in the anterior–posterior direction. For 73 tomotherapy plans, the gamma analysis results using the developed software were compared to those analyzed by medical physicists using a commercial software package.ResultsThe developed software successfully automated the gamma analysis for the tomotherapy delivery quality assurance. The gamma passing rate (GPR) calculated by the developed software was higher than that by the clinically used software by 3.0%, on average. While, for 1 of the 73 plans, the GPR by the manual gamma analysis was higher than 90% (pass/fail criteria), the gamma analysis using the developed software resulted in fail (GPR < 90%).ConclusionsThe use of automated and standardized gamma analysis software can improve both the clinical efficiency and veracity of the analysis results. Furthermore, the gamma analyses with various film scaling factors and setup uncertainties will provide clinically useful information for further investigations.

Funder

Yonsei University College of Medicine

National Research Foundation of Korea

Ministry of Trade, Industry and Energy

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3