Effect of Source-Detector Distance on the Measurement of Hemoglobin Using Near-Infrared Spectroscopy in Breast Cancer

Author:

Hayashi Maho1ORCID,Yoshizawa Nobuko1,Ueda Yukio1,Mimura Tetsuya2,Ohmae Etsuko2,Yoshimoto Kenji2,Wada Hiroko2,Nasu Hatusko1,Ogura Hiroyuki3,Sakahara Harumi1

Affiliation:

1. Department of Diagnostic Radiology & Nuclear Medicine, Hamamatsu University School of Medicine, Hamamatsu, Shizuoka, Japan

2. Central Research Laboratory, Hamamatsu Photonics K.K., Hamamatsu, Shizuoka, Japan

3. First Department of Surgery, Hamamatsu University School of Medicine, Hamamatsu, Shizuoka, Japan

Abstract

We measured total hemoglobin concentrations in breast tumors by near-infrared time-resolved spectroscopy. Muscles interfere with measurement when the probe is close to the chest wall. Since the target area of measurement depends on the distance between the light source and probe detector, we inferred that this issue could be solved by reducing the source-detector distance. The purpose of this study was to examine the effects of the source-detector distance on the measurement of total hemoglobin concentration in the breast. We examined 26 patients with breast tumors. Total hemoglobin concentration was measured in tumors and the contralateral normal breasts at source-detector distances of 20 and 30 mm. The difference in total hemoglobin concentration between each tumor and the contralateral breast was calculated. The normal breast total hemoglobin concentration was significantly smaller for the source-detector distance of 20 mm than for the source-detector distance of 30 mm. Differences in source-detector distance did not significantly affect tumor total hemoglobin. The difference in total hemoglobin concentration between the tumor and the contralateral breast obtained at the source-detector distance of 20 mm was significantly higher than that obtained at the source-detector distance of 30 mm. From these results, we considered that measurement with a source-detector distance of 20 mm is less affected by the chest wall than with a source-detector distance of 30 mm and that the difference in total hemoglobin concentration between the tumor and the contralateral breast at a source-detector distance of 20 mm can better reflect the net total hemoglobin concentrations of the breast tumors. In conclusion, using a probe with a source-detector distance of 20 mm can more accurately evaluate the total hemoglobin concentration in breast tumors.

Funder

Japan Society for the Promotion of Science

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3