The Measurement of the Surface Dose in Regular and Small Radiation Therapy Fields Using Cherenkov Imaging

Author:

Li Yi123ORCID,Liu HongJun14,Huang Nan1,Wang Zhaolu1,Zhang Chunmin2

Affiliation:

1. State Key Laboratory of Transient Optics and Photonics, Xi’an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi’an 710119, China

2. School of Physics, Xi’an Jiaotong University, Xi’an 710049, China

3. University of Chinese Academy of Sciences, Beijing 100049, China

4. Collaborative Innovation Center of Extreme Optics, Shanxi University, Taiyuan 030006, China

Abstract

Purpose: The aim of this study is to measure the output factor (OF) and profile of surface dose in regular and small radiation therapy fields using Cherenkov imaging (CI). Methods: A medical linear accelerator (linac) was employed to generate radiation fields, including regular open photon field (ROPF), regular wedge photon field (RWPF), regular electron field (REF) and small photon field (SPF). The photon beams consisted of two filter modes including flattening filter (FF) and flattening filter free (FFF). All fields were delivered to a solid water phantom. Cherenkov light was captured using a charge-coupled device system during phantom irradiation. The OF and profile of surface dose measured by CI were compared with those determined by film measurement, ionization chamber measurement and treatment planning system calculation in order to examine the feasibility of measuring surface dose OF and profile using CI. Results: The discrepancy between surface dose OF measured by CI and that determined by other methods is less than 6% in ROPFs with size less than 10 × 10 cm2, REFs with size less than 10 × 10 cm2, and SPFs except for 1 × 1 cm2 field. In the flat profile region, the discrepancy between surface dose profile measured by CI and that determined by other methods is less than 4% in REFs and less than 3% in ROPFs, RWPFs, and SPFs except for 1 × 1 cm2 field. The discrepancy of the surface dose profile is in compliance with the recommendation by IAEA TRS 430 reports. The discrepancy between field width measured by CI and that determined by film measurement is equal to or less than 2 mm, which is within the tolerance recommend by the guidelines of linac quality assurance in regular open FF photon fields, SPFs, and REFs with cone size of 10 × 10 cm2 in area. Conclusion: CI can be used to quantitatively measure the OF and profile of surface dose. It is feasible to use CI to measure the surface dose profile and field width in regular open FF photon fields and SPFs except for 1 × 1 cm2 field.

Funder

Qingdao National Laboratory for Marine Science and Technology of China

the Major International (Regional) Joint Research Project of National Natural Science Foundation of China

National High Technology Research and Development Program of China

the Key Program of National Natural Science Foundation of China

the National Natural Science Foundation of China

Publisher

SAGE Publications

Subject

Cancer Research,Oncology