Dosimetric Effect of Target Position Accuracy on Single-Isocenter Multiple Liver Metastases SBRT

Author:

Zhang Yun123ORCID,Niu Shanzhou4,Yuan Jun5,Wang Xiaoping123,Gong Changfei123ORCID,Tang Chunbo5ORCID

Affiliation:

1. Department of Radiation Oncology, Jiangxi Cancer Hospital, Nanchang, Jiangxi, PR China

2. The Second Affiliated Hospital of Nanchang Medical College, Nanchang, Jiangxi, PR China

3. Jiangxi Clinical Research Center for Cancer, JXHC Key Laboratory of Tumor Microenvironment and Immunoregulation (Jiangxi Cancer Hospital), Nanchang, Jiangxi, PR China

4. Ganzhou Key Laboratory of Computational Imaging, Gannan Normal University, Ganzhou, Jiangxi Province, China

5. Department of Oncology, First Affiliated Hospital of Gannan Medical University, Ganzhou, China

Abstract

Purpose: To evaluate the dosimetric effects of intrafraction baseline shifts combined with rotational errors on Four-dimensional computed tomography-guided stereotactic body radiotherapy for multiple liver metastases (MLMs). Methods: A total of 10 patients with MLM (2 or 3 lesions) were selected for this retrospective study. Baseline shift errors of 0.5, 1.0, and 2.0 mm; and rotational errors of 0.5°, 1°, and 1.5°, were simulated about all axes. All of the baseline shifts and rotation errors were simulated around the planned isocenter using a matrix transformation of 6° of freedom. The coverage degradation of baseline shifts and rotational errors were analyzed according to the dose to 95% of the planning target volume (D95) and the volume covered by 95% of the prescribed dose (V95), and related changes in gross tumor volume were also analyzed. Results: At the rotation error of 0.5° and the baseline offset of less than 0.5 mm, the D95 and V95 values of all targets were >95%. For rotational errors of 1.0° (combined with all baseline shift errors), 36.3% of targets had D95 and V95 values of <95%. Coverage worsened substantially when the baseline shift errors were increased to 1.0 mm. D95 and V95 values were >95% for about 77.3% of the targets. Only 11.4% of the D95 and V95 values were >95% when the baseline shift errors were increased to 2.0 mm. When the rotational error was increased to 1.5° and baseline shift errors increased to 1.0 mm, the D95 and V95 values were >95% in only 3 cases. Conclusions: The multivariate regression model analysis in this study showed that the coverage of the target decreased further with reduced target volume, increasing the baseline drift, the rotation error, and the distance to the target.

Funder

The "Five-level Progressive" talent cultivation project of Jiangxi Cancer Hospital & Institute

Jiangxi Double Thousand Plan

Joint Fund of the National Natural Science Foundation of China and the Karst Science Research Center of Guizhou Province

Science and Technology Program of Jiangxi Province

Publisher

SAGE Publications