Dosimetric Evaluation of the QFix kVueTM Calypso Couch Top

Author:

Hou Lingtong1,Zhang Huiqin2,Sun Xiaomei3,Liu Qianqian1,Chen Tingfeng1,Liu Yong1,Jiang Xiaodong2,Yao Shengyu1ORCID

Affiliation:

1. Department of Radiation Oncology, Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China

2. Department of Oncology, The Affiliated Lianyungang Hospital of Xuzhou Medical University, Lianyungang, China

3. Department of Gynecology, Obstetrics and Gynecology Hospital, Fudan University, Shanghai, China

Abstract

Purpose: To evaluate the dosimetric accuracy of the default couch model of the QFix kVueTM Calypso couch top in the treatment planning system. Methods: With the gantry 180°, field size 20 × 20 cm, 6 MV, we measured the depth dose, off-axis dose, and dose plane of different depths in the phantom with the couch rails in and out, respectively. Isocenter doses at different angles were also obtained. The results were compared to the doses calculated using the default couch top model and the real scanned couch top model. Then we revised the default model according to the measured results. Results: With “Rails In,” the depth dose, off-axis dose, and dose plane of the default couch top model had a big difference with the dose of the real scanned couch top model and the measured result. The dose of the real scanned couch top model was much closer to the measured result, but in the region of the rail edge, the difference was still significant. With “Rails Out,” there was a minor difference between the measured result, the dose of the default couch top model and the real scanned couch top model. The difference between the measurement and the default couch top model became very small after being revised. Conclusions: It is better to avoid the beam angle passing through the couch rails in treatment plans, or you should revise the parameter of the QFix kVueTM Calypso couch top model based on the measured results, and verify the treatment plan before clinical practice.

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Optimization of TomoTherapy couch density in the RayStation treatment-planning system;Journal of the Korean Physical Society;2023-11-14

2. Implementation of the Calypso system: a commissioning experience;Reports of Practical Oncology and Radiotherapy;2023-06-26

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3