Nondeformed Ultrasound Image Production Method for Ultrasound-Guided Radiotherapy

Author:

Li Qixuan1234ORCID,Zhang Fan1234,Xi Qianyi1234ORCID,Jiao Zhuqing15,Ni Xinye234

Affiliation:

1. School of Microelectronics and Control Engineering, Changzhou University, Changzhou, China

2. Department of Radiotherapy, The Affiliated Changzhou No.2 People's Hospital of Nanjing Medical University, Changzhou, China

3. Jiangsu Province Engineering Research Center of Medical Physics, Changzhou, China

4. Center for Medical Physics, Nanjing Medical University, Changzhou, China

5. School of Computer and Artificial Intelligence, Changzhou University, Changzhou, China

Abstract

Purpose: During ultrasound (US)-guided radiotherapy, the tissue is deformed by probe pressure, and the US image is limited by changes in tissue and organ position and geometry when the US image is aligned with computed tomography (CT) image, leading to poor alignment. Accordingly, a pixel displacement-based nondeformed US image production method is proposed. Methods: The correction of US image deformation is achieved by calculating the pixel displacement of an image. The positioning CT image (CTstd) is used as the gold standard. The deformed US image (USdef) is inputted into the Harris algorithm to extract corner points for selecting feature points, and the displacement of adjacent pixels of feature points in the US video stream is calculated using the Lucas-Kanade optical flow algorithm. The moving least squares algorithm is used to correct USdef globally and locally in accordance with image pixel displacement to generate a nondeformed US image (USrev). In addition, USdef and USrev were separately aligned with CTstd to evaluate the improvement of alignment accuracy through deformation correction. Results: In the phantom experiment, the overall and local average correction errors of the US image under the optimal probe pressure were 1.0944 and 0.7388 mm, respectively, and the registration accuracy of USdef and USrev with CTstd was 0.6764 and 0.9016, respectively. During the volunteer experiment, the correction error of all 12 patients’ data ranged from −1.7525 to 1.5685 mm, with a mean absolute error of 0.8612 mm. The improvement range of US and CT registration accuracy, before and after image deformation correction in the 12 patients evaluated by a normalized correlation coefficient, was 0.1232 to 0.2476. Conclusion: The pixel displacement-based deformation correction method can solve the limitation imposed by image deformation on image alignment in US-guided radiotherapy. Compared with USdef, the alignment results of USrev with CT were better.

Funder

the Social Development Project of Jiangsu Key R&D Plan

the Surface project of Jiangsu Provincial Health Commission

Publisher

SAGE Publications

Subject

Cancer Research,Oncology