Evaluation of VMAT Planning Strategies for Prostate Patients with Bilateral Hip Prosthesis

Author:

To David1ORCID,Xhaferllari Ilma1,Liu Manju1,Liang Jian2,Knill Cory3,Nandalur Sirisha1,Gustafson Gary1,Lack Danielle1

Affiliation:

1. Beaumont Health, Troy, MI, USA

2. Beaumont Health, Royal Oak, MI, USA

3. Beaumont Health, Dearborn, MI, USA

Abstract

Purpose: In this study, we investigate linac volumetric-modulated arc therapy (VMAT) planning strategies for bilateral hip prostheses prostate patients with respect to plan quality and deliverability, while limiting entrance dose to the prostheses. Methods: Three VMAT plans were retrospectively created for 20 patients: (1) partial arcs (PA), (2) 2 full arcs optimized with 500 cGy max prostheses dose (MD), and (3) 2 full arcs optimized with max dose-volume histogram (DVH) constraint of 500 cGy to 10% prostheses volume (MDVH). PA techniques contained 6 PA with beam angles that avoid entering each prosthesis. For each patient, other than prostheses constraints, the same Pinnacle VMAT optimization objectives were used. Plans were normalized with PTV D95% = 79.2 Gy prescription dose. Organ-at-risk DVH metrics, monitor units (MUs), conformality, gradient, and homogeneity indices were evaluated for each plan. Mean entrance prosthesis dose was determined in Pinnacle by converting each arc into static beams and utilizing only control points traversing each prosthesis. Plan deliverability was evaluated with SunNuclear ArcCheck measurements (gamma criteria 3%/2 mm) on an Elekta machine. Results: MD and MDVH had similar dosimetric quality, both improved DVH metrics for rectum and bladder compared to PA. Plan complexities among all plans were similar (average MUs: 441-518). Conformality, homogeneity, and gradient indices were significantly improved in MD and MDVH versus PA ( P < .001). Gamma pass rates for MD (99.0 ± 1.2%) and MDVH (99.2 ± 0.99%) were comparable. A significant difference over PA was observed (96.8 ± 1.6%, P < .001). Field-by-field analysis demonstrated 12/20 PA plans resulted in fields with pass rates <95% versus 1/20 plans for MD and none for MDVH. Cumulative mean entrance doses to each prosthesis were 62.9 ± 17.7 cGy for MD plans and 83.4 ± 27.5 cGy for MDVH plans. Conclusion: MD and MDVH plans had improved dosimetric quality and deliverability over PA plans with minimal entrance doses (∼1% of prescription) to each prosthesis and are an improved alternative for bilateral prostheses prostate patients.

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3