Fast MCsquare-Based Independent Dose Verification Platform for Pencil Beam Scanning Proton Therapy

Author:

Liu Chunbo12,Ho Meng Wei23,Park Jiyeon23,Hsi Wen Chien23,Liang Xiaoying23,Li Zuofeng23,Song Yuntao14,Feng Hansheng4,Zhang Yawei23ORCID

Affiliation:

1. School of Physical Sciences, University of Science and Technology of China, Hefei, Anhui, China

2. University of Florida Health Proton Therapy Institute, Jacksonville, FL, USA

3. Department of Radiation Oncology, University of Florida, Gainesville, FL, USA

4. Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei, Anhui, China

Abstract

Purpose: To commission MCsquare (a multi-cores CPU-based dose calculation engine) for pencil beam scanning (PBS) proton therapy, integrate it into RayStation treatment plan system (TPS) to create a dedicated platform for fast independent dose verification. Method: A MCsquare-based independent dose verification platform (MC2InRS) was developed to realize automatic dose re-calculation for clinical use, including data preparation, dose calculation, 2D/3D gamma analysis. MCsquare was commissioned based on in-air lateral dose profiles, integrated depth dose, and the absolute dose of different beam energies for Proteus®ONE. MC2InRS was validated with measurement data using various targets and depths in a water phantom. This study also investigated 15 clinical cases to demonstrate the feasibility and effectiveness of MC2InRS platform in clinic practice. Results: Between simulation and measurement, the distal range differences at 80% (R80) and 20% (R20) dose levels for each energy were below 0.05 mm, and 0.1 mm, respectively, and the absolute dose differences were below 0.5%. 29 out of 36 QA planes reached a 100% gamma passing rate (GPR) for 2%/2mm criteria, and a minimum of 98.3% gamma was obtained in water phantom between simulation and measurement. For the 15 clinical cases investigated, the average 2D GPR (2%/2mm) was 95.4%, 99.3% for MCsquare vs. measurement, MCsquare vs. TPS, respectively. The average 3D GPR (2%/2mm) was 98.9%, 95.3% for MCsquare vs. TPS in water, and computed tomography (CT), respectively. Conclusion: MC2InRS, a fast, independent dose verification platform, has been developed to perform dose verification with high accuracy and efficiency for Pencil Bream Scanning (PBS). Its potential to be applied in routine clinical practice has also been discussed.

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3