Optimizing effects on airway pressure and minute volume during closed endotracheal suctioning: a simulated lung model

Author:

Jung Fang123,Chou Shang-Shing P2,Yang Shih-Hsing1ORCID,Lin Jau-Chen1,Jow Guey-Mei4

Affiliation:

1. Department of Respiratory Therapy, Fu Jen Catholic University

2. Department of Chemistry, Fu Jen Catholic University

3. Graduate Institute of Biomedical and Pharmaceutical Science, Fu Jen Catholic University

4. School of Medicine, Fu Jen Catholic University

Abstract

A closed suction system is used to remove endotracheal secretions without interrupting the patient’s ventilation. Closed suctioning may reduce adverse effects associated with suctioning with, for example, decreased clinical signs of hypoxemia and limited environmental, personnel, and patient contamination. However, it is not clear whether ventilation is maintained during the procedure. We aimed to determine the effects of endotracheal tube (ETT) size, suction catheter (SC) size, and SC length in the ETT on ventilation parameters measured during suction. Suction was performed on a test lung, ventilated with either volume-controlled continuous mandatory ventilation (VC-CMV) or pressure-controlled continuous mandatory ventilation (PC-CMV) using ETT sizes of 6.0–8.5 mm paired with SC sizes of 8–16 French gauge (Fr = 0.33 mm). Airway resistance ( Raw), peak inspiratory pressure (PIP), positive end-expiratory pressure (PEEP), and expiratory minute volume ( Vexp) were recorded for each ventilation episode by a HAMILTON-G5 ventilator. Here, Raw was considerably increased by insertion of the SC into the ETT. This Raw effect altered the PIP and Vexp. PIP was increased in VC-CMV because the ventilation area of the ETT was reduced, and Vexp was decreased in PC-CMV in relation to the size of the SC. PEEP decreased with application of the 16 Fr SC and 30 L/min flow rate in VC-CMV. We conclude that airway pressure and minute volume are not maintained during closed endotracheal suctioning with VC-CMV and PC-CMV, respectively. The degree of interference to ventilation is affected through selection of appropriate SC size and ventilation settings.

Publisher

SAGE Publications

Subject

Computer Graphics and Computer-Aided Design,Modelling and Simulation,Software

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3