Analysis of Failed Ankle Arthroplasty Components

Author:

Currier Barbara H.1ORCID,Hecht Paul J.2,Nunley James A.3,Mayor Michael B.1,Currier John H.1,Van Citters Douglas W.1

Affiliation:

1. Thayer School of Engineering, Dartmouth College, Hanover, NH, USA

2. Dartmouth Hitchcock Medical Center, Lebanon, NH, USA

3. Duke University, School of Medicine, Durham, NC, USA

Abstract

Background: Although advances in joint-replacement technology have made total ankle arthroplasty a viable treatment for end-stage arthritis, revision rates for ankle replacements are higher than in hip or knee replacements. The questions asked in this study were what can retrieved ankle devices demonstrate about ankle arthroplasty failures and how can understanding the causes of these failures inform clinical decisions for current and future ankle arthroplasty patients? Methods: An IRB-approved retrieval laboratory received retrieved components and surgeon-supplied reason for revision from 70 total-ankles (7 designs, including 5 currently marketed designs) from 2002 to 2018. These retrievals were rated for clinical wear and damage. Metal components were rated by method and effectiveness of fixation. Polyethylene inserts received by the laboratory 6 months or less after retrieval (n = 45) were analyzed for oxidation using Fourier transform infrared spectroscopy. Statistical analysis was performed using IBM SPSS, version 22. Results: The ankle implants were retrieved most commonly for loosening and polyethylene fracture. Loosening occurred more frequently in fixed-bearing designs (n = 18) than in the mobile-bearing designs (n = 4) and after shorter in vivo time (mean in vivo time to retrieval for loosening: fixed bearing 3.2 ± 2.1 years, mobile bearing 9.7 ± 4.5 years). Gamma-sterilized ankle inserts oxidized at a higher rate than non-gamma (EtO or gas-plasma) sterilized ankle inserts (gamma 0.29 ± 0.22/y, non-gamma 0.07 ± 0.05/y, mean difference=0.215, 95% CI 0.128-0.303, P < .001). The presence of clinical fatigue (cracking and/or delamination) of the polyethylene insert correlated with measured oxidation (Spearman rho = 0.685, P < .001). Nine inserts, all gamma-sterilized, fractured in vivo. Conclusions: This study suggests that loosening could be more of a problem in fixed-bearing devices than in mobile bearing devices. Gamma-sterilized polyethylene inserts were found to suffer fatigue damage or fracture in vivo, resulting in the need for revision. Retrieval analysis can provide insight into implant-related reasons for revision, with the goal of understanding the implant-related causes of these failures, informing future ankle design and clinical decisions for current and future ankle arthroplasty patients. Level of Evidence: Level III, comparative series.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 35 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3