Early Radiographic and Clinical Outcomes of a Novel, Fixed-Bearing Fourth-Generation Total Ankle Replacement System

Author:

Kim Jaeyoung1ORCID,Rajan Lavan1ORCID,Bitar Rogerio1ORCID,Caolo Kristin1,Fuller Robert1,Henry Jensen1ORCID,Deland Jonathan1,Ellis Scott J.1ORCID,Demetracopoulos Constantine1

Affiliation:

1. Department of Orthopaedic Surgery, Hospital for Special Surgery, New York, NY, USA

Abstract

Background: The Cadence Total Ankle System is a 2-component, fixed-bearing fourth-generation total ankle arthroplasty (TAA) system that was introduced for clinical use in 2016. The purpose of this study was to report non–inventor, non–industry funded survivorship, radiographic and clinical outcomes, and early complications following use of this implant at a minimum of 2 years. Methods: This single-center retrospective study included patients who underwent TAA by 2 surgeons with this novel fixed-bearing system between January 2017 and September 2018. Forty-eight patients were evaluated at an average of 33.6 months. Radiographic outcomes included preoperative and postoperative tibiotalar angle on anteroposterior radiographs of the ankle, sagittal tibial angle (STA) on lateral radiographs of the ankle, and periprosthetic lucency formation and location. Revision and reoperation data were collected, and patient-reported outcomes were assessed using Patient Reported Outcomes Measurement Information System (PROMIS). Subgroup analysis assessed associations between preoperative deformity, postoperative implant alignment, PROMIS scores, and periprosthetic lucency formation. Results: Survivorship of implant was 93.7%, with 3 revisions, 1 due to infection and 2 due to loosening of the implant (1 tibial and 1 talar component). Three patients had reoperations (6.3%): 2 for superficial infection and 1 for gutter debridement due to medial gutter impingement. Fifteen patients (35.8%) developed periprosthetic lucencies, all on the tibial side. PROMIS scores improved after surgery in all domains except Depression. Patients with significant postoperative periprosthetic lucency had worse postoperative PROMIS Physical function scores than patients without lucency ( P < .05). Conclusion: This study demonstrated excellent minimum 2-year clinical and radiographic outcomes and low revision and reoperation rates of this new fourth-generation TAA system. Future studies with longer follow-up, especially on patients with periprosthetic lucency, are necessary to investigate the long-term complications and understand the long-term functional and radiographic outcomes of this implant.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3