Motion at the Tibial and Polyethylene Component Interface in a Mobile-Bearing Total Ankle Replacement

Author:

Lundeen Gregory A.1,Clanton Thomas O.2,Dunaway Linda J.1,Lu Minggen3

Affiliation:

1. Reno Orthopaedic Clinic, Reno, NV, USA

2. Foot and Ankle Sports Medicine, The Steadman Clinic, Vail, CO, USA

3. School of Community Health Sciences, University of Nevada, Reno, NV, USA

Abstract

Background: Normal biomechanics of the ankle joint includes sagittal as well as axial rotation. Current understanding of mobile-bearing motion at the tibial-polyethylene interface in total ankle arthroplasty (TAA) is limited to anterior-posterior (AP) motion of the polyethylene component. The purpose of our study was to define the motion of the polyethylene component in relation to the tibial component in a mobile-bearing TAA in both the sagittal and axial planes in postoperative patients. Methods: Patients who were a minimum of 12 months postoperative from a third-generation mobile-bearing TAA were identified. AP images were saved at maximum internal and external rotation, and the lateral images were saved in maximum plantarflexion and dorsiflexion. Sagittal range of motion and AP translation of the polyethylene component were measured from the lateral images. Axial rotation was determined by measuring the relative position of the 2 wires within the polyethylene component on AP internal and external rotation imaging. This relationship was compared to a table developed from fluoroscopic images taken at standardized degrees of axial rotation of a nonimplanted polyethylene with the associated length relationship of the 2 imbedded wires. Sixteen patients were included in this investigation, 9 (56%) were male and average age was 68 (range, 49-80) years. Time from surgery averaged 25 (range, 12-38) months. Results: Total sagittal range of motion averaged 23±9 (range, 9-33) degrees. Axial motion for total internal and external rotation of the polyethylene component on the tibial component averaged 6±5 (range, 0-18) degrees. AP translation of the polyethylene component relative to the tibial component averaged 1±1 (range, 0-3) mm. There was no relationship between axial rotation or AP translation of the polyethylene component and ankle joint range of motion ( P > .05). Conclusion: To our knowledge, this is the first investigation to measure axial and sagittal motion of the polyethylene component at the tibial implant interface in patients following a mobile-bearing TAA. Based on outcome scores and range-of-motion measurements, we believe the patients in this study are a representative cross section of subjects compared to other TAA research results. The results from this investigation indicate the potential for a mobile-bearing TAA to fall within the parameters of normal polyaxial ankle motion. The multiplanar articulation in a mobile-bearing TAA may reduce excessively high peak pressures during the complex dynamic tibial and talar motion, which may have a positive influence on gait pattern, polyethylene wear, and implant longevity. Level of Evidence: Level IV, case series.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3