Biomechanical Comparison of Syndesmosis Fixation with 3.5- and 4.5-millimeter Stainless Steel Screws

Author:

Thompson Michael C.1,Gesink Dirk S.1

Affiliation:

1. Department of Orthopaedic Surgery and Rehabilitation University of Nebraska Medical Center 981080 Nebraska Medical Center Omaha, NE 68198-1080 Fax (402) 559-5511

Abstract

Although most authors recommend either 3.5-mm or 4.5-mm cortical screws for syndesmosis fixation, the optimum screw size has yet to be defined. The present study was designed to biomechanically compare syndesmosis fixation with 3.5-mm and 4.5-mm stainless steel screws. Simulated pronation external rotation ankle injuries were created in twelve paired, fresh-frozen cadaveric leg specimens. One limb from each pair received a 3.5-mm tricor-tical stainless steel screw for syndesmosis fixation (group I), while the contralateral specimen was stabilized using a 4.5-mm screw (group II). Sub-maximal axial ramp (0 to 1200 N) and external rotation/torsional ramp (0 to 5 N-m) loading was performed on each specimen prior to ligament division, following ligament division and following syndesmosis fixation. Axial fatigue testing was then performed at 1.5 Hz for a total of 100,000 cycles (0 to 900 N), and each specimen was subsequently tested to failure in external rotation. Ligament division resulted in syndesmosis widening (p<0.001) and reduced stiffness (p<0.001) during torsional ramp loading. Subsequent syndesmosis screw placement reduced syndesmosis widening (p<0.05) and increased stiffness (p<0.05). Following screw fixation, however, widening remained greater (p<0.005) and stiffness less (p<0.001) than pre-injury levels. No differences between groups I and II were observed during submaximal testing. In external rotation to failure testing, group I failed at a greater angle (38.9° ± 4.1° vs. 32.0° ± 3.8° in group II; p<0.05). Failure torque was slightly higher in group I; however, the difference was not statistically significant (17.8 ± 2.0 N-m vs. 14.3 ± 2.6 N-m in group II; p=0.082). Five specimens in group I failed by screw pullout and five specimens in group II failed by fibula fracture (p=0.061). The present results suggest that there is no biomechanical advantage of a 4.5-mm screw over a 3.5-mm in fixation of the syndesmosis.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 121 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3