Affiliation:
1. Department of Orthopedic Surgery, NYU Langone Health, New York, NY, USA
Abstract
Study Design Retrospective Cohort Study Objectives: Robot-assisted spine surgery (RASS) is a rapidly evolving technique with potential benefits for improving surgical outcomes. A number of studies on RASS learning curve have focused on early iterations of the Mazor robot. Limited research exists on the learning curve associated with using the Globus Medical ExcelsiusGPS® system. In this retrospective study, we aimed to evaluate the learning curve of RASS using the ExcelsiusGPS® system at a single institution. Methods A total of 95 patients (541 screws) who underwent RASS between 2021 and 2022 were included. Variables including operative time, robot registration time, screw placement time, fluoroscopy utilization, and complications were analyzed. Statistical analysis was performed using descriptive statistics and two-sample t-tests. Results The average operative time significantly decreased after the first 14 cases, indicating a learning curve. However, no significant improvement was observed in robot registration time. Notably, screw placement time significantly improved after approximately 13 cases. When controlling for the number of levels fused, the trends remained consistent. Conclusions Our study confirmed the presence of a learning curve in RASS using the ExcelsiusGPS® system and demonstrated rapid proficiency development. Our findings highlight the relatively quick learning curve of 1 RASS system.