Fatigue life estimation of dental implants using a combination of the finite element method and traditional fatigue criteria

Author:

Hernandez Bruno A1ORCID,Freitas João PO2,Capello Sousa Edson A2

Affiliation:

1. Department of Mechanics, Centre for Simulation in Bioengineering, Biomechanics and Biomaterials (CS3B), School of Engineering, Campus of Guaratinguetá, São Paulo State University (UNESP), Guaratinguetá, São Paulo State, Brazil

2. Department of Mechanical Engineering, Centre for Simulation in Bioengineering, Biomechanics and Biomaterials (CS3B), School of Engineering, Campus of Bauru, São Paulo State University (UNESP), Bauru, São Paulo State, Brazil

Abstract

Failure by fatigue can be sudden and catastrophic. Therefore, ensuring that dental implants, which are under constant cyclic loading, do not fail to fatigue is imperative. The majority of the studies about the topic only performed in vitro tests, which are expensive and time-consuming. The Finite Element (FE) method is less costly and it allows the simulation of several different loading scenarios. Nonetheless, there are only a few studies analysing fatigue in dental prostheses using FE models, and the few available did not include all the relevant parameters, such as geometry effect, surface finishing, etc. Therefore, this study aimed to analyse the fatigue behaviour of a single-unit dental implant with two screws using a combination of the numerical results and the traditional fatigue criteria – a combination that was not yet fully and correctly explored. A finite element model comprising a single implant, one abutment, one abutment screw, one fixation screw and one prosthetic crown was developed. Material properties were assigned based on literature data. A 100  N load was applied to mimic the mastication forces and fatigue analysis was conducted using the Gerber, Goodman and Soderberg fatigue criteria. The fatigue analysis demonstrated that the abutment screw could fail in less than 1 year, depending on the criteria, while the fixation screw exhibits an infinite life. The results illustrated the importance of analysing the fatigue behaviour of dental implants and highlighted the potential of finite element models to simulate the biomechanical behaviour of dental implants.

Funder

Coordenacao de Aperfeicoamento de Pessoal de Nivel Superior

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine