Distal humerus cortical strains following total elbow arthroplasty

Author:

Craik Johnathan D12,Laffer Caroline HL2,Richards Simon W3,Walsh Sean P4,Evans Sam L1

Affiliation:

1. Department of Engineering, Cardiff University, Epsom, UK

2. Department of Trauma and Orthopaedics, Epsom Hospital, Epsom, UK

3. Department of Orthopaedics, Poole Hospital NHS Foundation Trust, Poole, UK

4. Department of Orthopaedics, Dorset County Hospital NHS Foundation Trust, Dorchester, UK

Abstract

The aseptic loosening of total elbow replacements is a serious complication resulting in significant patient morbidity. It is thought to occur secondary to stress shielding of the distal humeral cortex by the stiff stem of the implant. Some total elbow prostheses incorporate an anterior flange intended to improve implant stability and peri-articular load transfer in an attempt to reduce this effect. However, few studies have directly assessed the changes in cortical strains following total elbow arthroplasty or the biomechanical advantage of the anterior flange design. A regular and a long flange Coonrad–Morrey total elbow prosthesis were implanted into six Sawbone synthetic humeri. The constructs were subjected to physiological loads in axial compression (500 N), antero-posterior bending (50 N) and antero-posterior compression with condylar supports (300 N). Digital image correlation was used to measure the distal antero-lateral cortical strains and the results compared with those of whole Sawbones that had been tested in the same way. Significant stress shielding was demonstrated over the distal humeral cortex following prosthesis implantation during axial compression. In contrast, cortical strains increased following prosthesis implantation during antero-posterior compression with condylar supports. The increase in cortical strains following total elbow arthroplasty may help to maintain the integrity of the anterior cortex offering additional stability for implants with an anterior flange. These results are important for the development of future total elbow prosthesis designs and indicate that simulating the action of the forearm muscles is essential when evaluating changes in strain about the distal humerus in vitro.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3