Clinical failure analysis of contemporary ceramic-on-ceramic total hip replacements

Author:

Brandt Jan-M1,Gascoyne Trevor C1,Guenther Leah E1,Allen Andrew2,Hedden David R1,Turgeon Thomas R13,Bohm Eric R1

Affiliation:

1. Concordia Joint Replacement Group, Winnipeg, MB, Canada

2. School of Biomedical Engineering, Dalhousie University, Halifax, NS, Canada

3. Orthopaedic Innovation Centre, Inc., Winnipeg, MB, Canada

Abstract

The present study investigates the performance of ceramic-on-ceramic total hip replacements by combining a retrieval analysis with a survivorship analysis to elucidate mechanisms that led to clinical failure. Semiquantitative surface damage assessment, contact profilometry, contour measurements, and scanning electron microscopy were performed to characterize the types and quantify the extent of surface damage on the retrieved ceramic components. The implantation period was positively correlated with both damage scores of the femoral heads (R = 0.573, p < 0.001) and the acetabular cups (R = 0.592, p < 0.001). Increased maximal out-of-roundness values of the femoral heads correlated with both increased metal transfer damage score (R = 0.384, p = 0.023) and increased stripe damage score (R = 0.729, p ≤ 0.001) of the acetabular liners. The damage rate (damage score/year) for both the retrieved heads and acetabular liners was at least 2.2-fold greater at inclination angles of >45° than the damage rate at inclination angles of ≤45°. For the retrieved femoral heads only, the linear wear rate of 25.5 ± 21.3 µm/year at inclination angles of >45° was 6-fold greater than the linear wear rate of 4.2 ± 2.3 µm/year at inclination angles of ≤45°. Metal transfer on the ceramic bearing surface could possibly contribute to fluid-film starvation and, in combination with an increased inclination angle, may facilitate an adhesive wear mechanism associated with stripe surface damage. At our institution, the clinical survivorship of ceramic-on-ceramic total hip replacements was 98.9% (a total of 9 out of 815 patients were revised within 10 years after total hip arthroplasty) with revision as the end point, suggesting their safe use in younger patients.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com