Abstract
This paper proposes an inverse-model-based cuffless method for estimating blood pressure using a single photoplethysmography sensor. The proposed method, which is based on the relationship between blood pressure and the features of pulse waves, employs an inverse estimation and uses the blood pressure as the explanatory variable. Using this method, the blood pressure can be estimated with high accuracy even in situations where the pulse wave features are scattered, as the method uses the dynamic signal-to-noise ratio of the Taguchi method. In order to verify the effectiveness of the proposed method, we employed it to measure the systolic blood pressure. It could be confirmed that the estimation accuracy of the proposed method is higher than that of similar methods.