Early failure of a highly cross-linked polyethylene inlay after total hip arthroplasty probably due to insufficient irradiation

Author:

Wahl Peter1ORCID,Mossu-Haas Christine2,Dommann-Scherrer Corina3,Wei Kongchang4,Eschbach Lukas5,Gehr Peter6,Benninger Emanuel1

Affiliation:

1. Division of Orthopaedics and Traumatology, Cantonal Hospital Winterthur, Winterthur, Switzerland

2. Central Station Medical Group, Fribourg, Switzerland

3. Institute of Pathology, Cantonal Hospital Winterthur, Winterthur, Switzerland

4. Laboratory for Biomimetic Membranes and Textiles, Empa, Swiss Federal Laboratories for Materials Science and Technology, St. Gallen, Switzerland

5. RMS Foundation, Bettlach, Switzerland

6. Institute of Anatomy, University of Berne, Berne, Switzerland

Abstract

Highly cross-linked polyethylene (XLPE) is a major advance in total hip arthroplasty (THA), as it suffers from less wear and thus is associated with lower revision rates than standard ultra-high molecular weight polyethylene. Early failures are reported rarely, and associated with specific design or manufacturing issues. We report a case requiring early revision due to adverse reaction to polyethylene particles. Investigations identified insufficient irradiation as the most probable cause of failure. Here are reported the features of a clinical case with determination of the material properties of the retrieved XLPE liner and establishment of the appropriate calibration curves as reference. Periprosthetic joint infection could be ruled out with appropriate sampling as cause for the inflammatory periarticular tumour. Histology identified a massive macrophagic reaction to micrometric polyethylene particles. No component malposition was present, nor any third-body wear. The trans-vinylene index (TVI) indicated insufficient and potentially detrimental irradiation of the polyethylene, while gel content, crystallinity, melting temperature and oxidation index remained within expected ranges. Histologically proven failure of a XLPE THA liner was identifiable despite correct implantation of the components. The cause of failure most probably was an inadequate irradiation, as indicated by determination of the TVI. This case underscores the importance of histologic workup even in aseptic revisions, and of detailed analysis of retrievals. The calibration curves provided are essential for analysis of other retrievals.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3