Perforated H-beam implant can be used in femoral neck fracture

Author:

Jafarov Afgan1,Erbay Elibol Fatma Kübra2,Alizadeh Chingiz3,Gülşen Mahir4,Toğrul Emre4,Demir Teyfik5ORCID

Affiliation:

1. Department of Orthopedics and Traumatology Surgery, Modern Hospital, Baku, Azerbaijan

2. Department of Micro and Nanotechnology, TOBB University of Economics and Technology, Ankara, Turkey

3. Scientific Research Traumatology and Orthopedic Institute, Baku, Azerbaijan

4. Department of Orthopedics and Traumatology, Private Ortopedia Hospital, Adana, Turkey

5. Department of Mechanical Engineering, TOBB University of Economics and Technology, Ankara, Turkey

Abstract

This biomechanical study evaluated comparison osteosynthesis of the femoral neck fracture model by 3 cannulated screw implants and new perforated H beam implants under different loading conditions with 45 third generation right proximal femur bones. A standardized Pauwels Type 3 of the femoral neck fracture was performed in the femur models. For assessing the rigidity and strength of fixation methods, the proximal femur bones after their osteosynthesis were then mechanically tested in axial compression, and torsional and dynamic axial compression loading. To determine the structural advantage of the new implant system, perforated and nonperforated new implant systems were comparing about pull out performance. When loading the samples, photographs were taken continuously. The reference parameters were described and measured from unloaded and loaded photographs of the static and dynamic tests. There was no significant difference between stiffness values of two fixation methods under static and rotational loading. Under dynamic loading, the displacement of the superior point of femoral head at the fracture line showed a significant decrease between the new implant system and cannulated screws. Comparing the relative motion at the mid line of the fracture in femoral neck between groups, a significant increase was found in H Beam implant group. Perforated H beam implants have similar static and torsion properties with golden standard. Although there was significant difference under dynamic loading which simulate movement early after surgery, the patient was not allowed to move early after surgery in the clinical practice. Therefore, the differences due to the perforated “H” beam implant would not cause clinical insecurity. Therefore, it is assumed that the perforated “H” beam implant can be used for internal fixation as an alternative to cannulated screws in the treatment of instable femoral neck fracture.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3