Influence of stem type on material loss at the metal-on-metal pinnacle taper junction

Author:

Hothi Harry S1,Whittaker Robert K1,Meswania Jay M1,Blunn Gordon W1,Skinner John A1,Hart Alister J1

Affiliation:

1. Institute of Orthopaedics and Musculoskeletal Science, University College London and Royal National Orthopaedic Hospital, Stanmore, UK

Abstract

The clinical importance of material loss at the head–stem junction is unknown. Comparison of retrievals with different stem types can provide the opportunity to understand the importance of the taper junction. This was a case–control study involving 20 retrieved 36 mm metal-on-metal Pinnacle (DePuy) hips that were paired with either a Corail (n = 10) or S-ROM (n = 10) stem. The median head taper material loss rate for the Corail group was 0.238 (0.0002–2.178) mm3/year and was significantly greater than the S-ROM group (p = 0.042), which had a median material loss rate of 0.132 (0.015–0.518) mm3/year. The only significant difference between the groups was the stem taper roughness and length: this was rougher and shorter for the Corails. Long and smooth stem taper designs are preferred when used in conjunction with metal heads.