Analysis of failed Van Straten LPM proximal interphalangeal prostheses

Author:

Bone M. C.1,Cunningham J. L.2,Lord J.1,Giddins G.3,Field J.4,Joyce T. J.1

Affiliation:

1. School of Mechanical and Systems Engineering, Newcastle University, Newcastle upon Tyne, UK

2. Department of Mechanical Engineering, University of Bath, Bath, UK

3. Orthopaedic Department, Royal United Hospital, Bath, UK

4. Cheltenham General Hospital, Cheltenham, UK

Abstract

The two-piece Van Straten Leuwen Poeschmann Metal (LPM) prosthesis was intended for the proximal interphalangeal joints. However, revision rates of 29% after 19 months were reported, as well as massive osteolysis. Five failed LPM titanium–niobium coated cobalt chromium components were obtained, three distal and two proximal, and subjected to a forensic retrieval analysis. Components were analyzed using a Talysurf contacting profilometer, ZYGO noncontacting profilometer, and environmental-scanning electron microscope. All components were heavily worn. In some regions the titanium–niobium coating had been scratched and penetrated. Elsewhere this coating had been removed where there was minimal scratching, which may have been due to corrosion between the coating and substrate. The osteolysis reported clinically was likely to be linked to the wear debris from the failed titanium–niobium coating and substrate.

Publisher

SAGE Publications

Subject

Surgery

Reference21 articles.

1. Characterization of Zr-based hard coatings for medical implant applications

2. British Standards. BS 7251-4 orthopaedic joint prostheses: Specification for bearing surfaces of hip joint prostheses. London, British Standards Institution, 1997.

3. TWO TO FIVE YEAR FOLLOW-UP OF THE LPM CERAMIC COATED PROXIMAL INTERPHALANGEAL JOINT ARTHROPLASTY

4. Editorial

5. Review Article: Osteolysis After Total Knee Arthroplasty

Cited by 10 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Explant analysis of a chromium nitride coated metal-on-metal total wrist replacement: A case study;Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine;2024-09-10

2. Wrist joint replacements;Joint Replacement in the Human Body;2024

3. Proximal interphalangeal joint replacements;Joint Replacement in the Human Body;2024

4. Research Progress on Wear and Corrosion Resistance of Artificial Joint and Disc by Plasma Surface Modification;CHINA SURF ENG;2019

5. Failure of an ACCIS metal-on-metal hip resurfacing prosthesis: A case report;Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine;2017-11-04